PRO MAM ACCREDITATION FF


Share


Products Description

Overview
This Full Field phantom was designed to test the performance of a digital mammographic system by evaluating the system’s ability to image small structures similar to those found clinically: micro-calcifications, fibrous structures in ducts and tumor-like masses. It is similar to the Pro-MAM Accreditation phantom, the main difference being the size: the FF (Full Field) version is larger and covers the entire image detector, thus eliminating scatter.


Optical density in the reference point
Spatial resolution
Dynamic range – image contrast
Threshold contrast visibility
Effective radiation field
Film processing
Automatic Exposure Control

Technical data (can be modified to customer specifications) :
Dimensions: 311.2 x 190.5 x 41.3 mm
Simulates 42 mm compressed breast of average glandular/adipose composition (50%/50%)
Comfortable carrying case (optional)
wax Insert:
- Nylon fibrils diameters: 0.89, 0.75, 0.61, 0.54, 0.40 and 0.30 mm
- Microcalcifications: 0.32, 0.28, 0.23, 0.20, 0.17 and 0.14 mm Al2O3 specks
- Tumor-like masses: 1.00, 0.75, 0.50, 0.38, 0.25 and 0.20 mm thick

Product Features :
Complies with :
- IEC 61223-3-2
- ACR Mammography Accreditation Program Requirements
- ACR Stereotactic Breast Biopsy Accreditation Program

Certified :
- CE

The Manual :
- Provides detailed guidelines for carrying out each test, results assessment and registration
Note a very good correlation of Contrast Details Curves of the Pro-MAM Gold mk II phantom against the limiting values set by EUREF. This ensures unmatched evaluation of a mammography system performance according to requirements of European Protocol for quality assurance in Digital Mammography.

Request A Call Back

I would like to discuss about
  • Looking for Radiation Services
  • Know more about our Products
  • Explore Accessories

We are Professional Radiation Services Company

Hospitals and Institutes who deal with X-Ray and radioactive material and equipment's are responsible for ensuring that their radiation minimize harm and risk to Technicians, workers and Public which will enable you to comply with Atomic Energy Regulatory Board regulations.

  • Professionals - Our Scientific Staff Are Qualified In Fields Ranging From Medical Physics To Nuclear Technologies.
  • Responsibility - We Work Closely With Our Clients In Order To Provide An Honest Service For Optimal Results.
  • Availability - We Are A Team Ready To Work For You Using Our Purpose Built Lab, Training Room And Office.
  • Work With Us - Our Professional Recognition In The Field Of Radiation Protection Is The Silent Promoter Of Our Work Quality.